TOMEs
Blood center operational equipment software. Attributes come from Terumo's own public documentation for this product.
Attributes
Signals
Properties (MITRE EMB3D)
PID-2322 Device deploys an access enforcement/privilege mechanism PID-31 Application-level software is present and running on the device PID-332 Device includes authenticated services PID-41 Device exposes remote network services PID-4113 Device includes cryptographic functions for sensitive data, such as encryption or authentication
Attributes: component types
API Generic Application Library Local Service Operating System
Attributes: interface types
TCP
Attributes: assets
Log Data Patient Data Report Data Software Configuration User Data
Attributes: dataflows
TOMEs to Blood Bank Information System (bidirectional, donor eligibility and expiration date validation) TOMEs to Custom Reports and Procedure Data Export TOMEs to Information Management Systems and TOMEs-compatible Devices (bidirectional data transmission) TOMEs-connected Devices to TOMEs (device and procedure status, procedure data) Third-party Digital Scales to TOMEs
Matched threat intel
Top 20 of 24
Advisories in our coverage that share attributes with this product, strongest first: a named product or component, the same device type or lineage, then the most shared attributes. A shared attribute is a lead, not a finding. The status column shows only what Terumo has said.
| # | Advisory | Date | Shared | Terumo status |
|---|---|---|---|---|
| 1 | Default credentials across a line of diagnostic instruments | Dec 17, 2024 | Same property (MITRE EMB3D)PID-3329 shared attributes: Blood, Handles patient data, Wired network, Remote access +5 | No response from manufacturer |
| 2 | A hard-coded VNC password on a bioreactor, base score 9.8 | May 26, 2026 | Same property (MITRE EMB3D)PID-4119 shared attributes: Remote access, User accounts, Wired network, Device includes authenticated services (PID-332) +5 | No response from manufacturer |
| 3 | Stored passwords recoverable in a Siemens imaging platform | Feb 10, 2026 | Exposed to the same EMB3D threatTID-2216 shared attributes: User accounts, Role-based access, Remote access, Handles patient data +2 | No response from manufacturer |
| 4 | BD Pyxis products may still run on shared default operating system credentials | May 31, 2022 | Same property (MITRE EMB3D)PID-236 shared attributes: User accounts, Handles patient data, Wired network, Operating System +2 | No response from manufacturer |
| 5 | GE Healthcare imaging and ultrasound systems expose default service credentials | Dec 8, 2020 | Same property (MITRE EMB3D)PID-4117 shared attributes: User accounts, Remote access, Handles patient data, Generic Application +3 | No response from manufacturer |
| 6 | Philips Alice 6 through R8.0.3 lacks authentication and encryption between nodes | Mar 27, 2018 | Exposed to the same EMB3D threatTID-4116 shared attributes: Wired network, User accounts, Handles patient data, Generic Application +2 | No response from manufacturer |
| 7 | Philips Interoperability Solution XDS can expose LDAP system credentials in cleartext | Jun 24, 2021 | Exposed to the same EMB3D threatTID-4115 shared attributes: User accounts, Wired network, Device exposes remote network services (PID-41), Local Service +1 | No response from manufacturer |
| 8 | BD Synapsys sessions did not expire, risking changes to patient data | May 31, 2022 | Same property (MITRE EMB3D)PID-3324 shared attributes: User accounts, Handles patient data, Device includes authenticated services (PID-332), Patient Data | No response from manufacturer |
| 9 | Philips PageWriter TC cardiographs expose a hard-coded superuser password and unchecked input | Aug 16, 2018 | Exposed to the same EMB3D threatTID-3114 shared attributes: User accounts, Operating System, Generic Application, Software Configuration | No response from manufacturer |
| 10 | BD Pyxis dispensing and supply systems ship with hard-coded credentials across 24 products | Mar 3, 2022 | Exposed to the same EMB3D threatTID-3114 shared attributes: User accounts, Handles patient data, Generic Application, Patient Data | No response from manufacturer |
| 11 | Baxter Phoenix hemodialysis system sends treatment and prescription data in cleartext | Jun 18, 2020 | Same signals4 shared attributes: Handles patient data, Generic Application, TCP, Patient Data | No response from manufacturer |
| 12 | Medtronic CareLink 9790, 2090 and Encore 29901 programmers store patient data unencrypted | Dec 13, 2018 | Exposed to the same EMB3D threatTID-2213 shared attributes: Handles patient data, Patient Data, Report Data | No response from manufacturer |
| 13 | Hillrom Welch Allyn ELI resting electrocardiographs contain hard-coded passwords | Jun 16, 2022 | Exposed to the same EMB3D threatTID-3111 shared attribute: User accounts | No response from manufacturer |
| 14 | BridgeLink 26.9.0 closes the Mirth Connect flaws CISA flagged, plus eight inherited CVEs | Oct 1, 2026 | 10 shared attributesApplication-level software is present and running on the device (PID-31), Device exposes remote network services (PID-41), Device includes authenticated services (PID-332), Device includes cryptographic functions for sensitive data, such as encryption or authentication (PID-4113) +6 | No response from manufacturer |
| 15 | Compounding software that did not encrypt its database traffic by default | May 15, 2026 | 11 shared attributesHandles patient data, Wired network, Remote access, User accounts +7 | No response from manufacturer |
| 16 | Cepheid tells GeneXpert Dx 6.4a sites to switch off TLS 1.0 and TLS 1.1 | Mar 12, 2026 | 8 shared attributesWired network, Device includes cryptographic functions for sensitive data, such as encryption or authentication (PID-4113), Operating System, Local Service +4 | No response from manufacturer |
| 17 | Hamilton Medical T1 ventilator through 2.2.3 has hard-coded admin credentials | Feb 16, 2021 | 6 shared attributesUser accounts, File import, Device includes authenticated services (PID-332), Local Service +2 | No response from manufacturer |
| 18 | Philips IntelliBridge EC40 and EC80 hubs allow weak SSH ciphers | Nov 14, 2019 | 8 shared attributesRemote access, Handles patient data, Device exposes remote network services (PID-41), Device includes cryptographic functions for sensitive data, such as encryption or authentication (PID-4113) +4 | No response from manufacturer |
| 19 | Illumina Universal Copy Service on sequencers binds to every IP address and runs with excess privileges | Apr 27, 2023 | 7 shared attributesFile import, Device exposes remote network services (PID-41), Local Service, Operating System +3 | No response from manufacturer |
| 20 | Thermo Fisher Torrent Suite Dx flaw CVE-2026-41085 lets limited users gain administrator access | May 15, 2026 | 5 shared attributesUser accounts, Role-based access, Device includes authenticated services (PID-332), Generic Application +1 | No response from manufacturer |
Threats and weaknesses in matched advisories
MITRE EMB3D threats
TID-202 Exploitable System Network Stack Component TID-210 Device Vulnerabilities Unpatchable TID-221 Authentication Bypass By Message Replay TID-226 Device leaks security information in logs TID-302 Install Untrusted Application TID-310 Remotely Accessible Unauthenticated Services TID-311 Default Credentials TID-317 Predictable Cryptographic Key TID-320 SQL Injection TID-324 HTTP Direct Object Reference TID-326 Insecure Deserialization TID-327 Out of Bounds Memory Access TID-328 Hardcoded Credentials TID-329 Improper Password Storage TID-406 Unauthorized Messages or Connections TID-408 Unencrypted Sensitive Data Communication TID-411 Weak/Insecure Cryptographic Protocol
CWE weaknesses
CWE-1104 Use of Unmaintained Third Party Components CWE-1188 Insecure Default Initialization of Resource CWE-1392 Use of Default Credentials CWE-1395 Dependency on Vulnerable Third-Party Component CWE-200 Exposure of Sensitive Information to an Unauthorized Actor CWE-276 Incorrect Default Permissions CWE-284 Improper Access Control CWE-294 Authentication Bypass by Capture-replay CWE-319 Cleartext Transmission of Sensitive Information CWE-321 Use of Hard-coded Cryptographic Key CWE-326 Inadequate Encryption Strength CWE-522 Insufficiently Protected Credentials CWE-639 Authorization Bypass Through User-Controlled Key CWE-798 Use of Hard-coded Credentials CWE-916 Use of Password Hash With Insufficient Computational Effort