CISA published advisory ICSMA-20-170-03 (Update A) on June 18, 2020, for a cleartext transmission vulnerability in the Baxter Phoenix Hemodialysis Delivery System, versions 3.36 and 3.40. CVE-2020-12048 (CWE-319) carries a CVSS v3 base score of 7.5. Baxter reported it.
The Phoenix does not support data-in-transit encryption such as TLS when it sends treatment and prescription data to the Exalis dialysis data management tool. Anyone with access to that network segment can read the traffic. The vector is network, low complexity, no privileges, with the impact limited to confidentiality.
CISA's risk statement is narrow: an attacker with unauthorized network access could view sensitive data. Nothing in the advisory says the link can be used to alter a treatment. Prescription data in the clear is still PHI in the clear, and the same missing encryption on a link that carries settings would be a far worse finding.
Baxter's answer is implementation guidance and a product security bulletin, plus a reminder that users must assess the risks of integrating the device into an enterprise network and repeat that analysis when the network changes. Unencrypted device-to-server links belong on a segment where the only listener is the server.