CISA published advisory ICSMA-20-343-01 on December 8, 2020, covering two vulnerabilities across GE Healthcare imaging and ultrasound products. Both carry a CVSS v3 score of 9.8. Lior Bar Yosef and Elad Luz of CyberMDX reported them.

CVE-2020-25179 (CWE-497) means exposed or default credentials can be used to access or modify sensitive information. CVE-2020-25175 (CWE-523, Unprotected Transport of Credentials) means specific credentials may be exposed in transit over the network. Both vectors are network, low complexity, no privileges and no user interaction.

The product list is long. It spans LightSpeed, BrightSpeed, Discovery CT and Optima CT scanners, Signa and Brivo MR systems, Definium, Discovery XR and Brivo XR radiography, Discovery NM and Infinia nuclear medicine cameras, Innova interventional systems, AW and AWS workstations, and LOGIQ, EchoPAC and Image Vault ultrasound software.

The attacker first needs a foothold on the healthcare delivery organization's network. From there, access is comparable to GE's own remote service user privileges, which can expose a limited set of PHI or run arbitrary code that affects availability and allows manipulation of PHI.

GE says it has mitigations for specific products and releases and will proactively configure product firewall protection and change default passwords on impacted devices where possible. Details sit on the GE Healthcare Product Security Portal. A service account that is the same on every scanner in the fleet is a fleet-wide key, and the fix has to reach each box.