CISA published advisory ICSMA-23-117-01 on April 27, 2023, for the Universal Copy Service (UCS) that ships on Illumina instruments. Illumina reported the vulnerabilities.
CVE-2023-1968 (CWE-1327) scores CVSS v3 10.0. Instruments with UCS v2.x bind to an unrestricted IP address, so an unauthenticated actor could use UCS to listen on all IP addresses, including those capable of accepting remote communications. CVE-2023-1966 (CWE-250, CVSS 7.4) affects UCS v1.x and v2.x: the service runs with unnecessary privileges, and an unauthenticated actor could upload and execute code at the operating system level.
Affected software includes MiSeq Control Software 4.0, MiSeqDx Operating Software 4.0.1 and later, MiniSeq Control Software 2.0 and later, NextSeq 1000/2000 Control Software 1.4.1 and earlier, NextSeq 500/550 and 550Dx Control Software 4.0, NextSeq 550Dx Operating Software, NovaSeq 6000 Control Software 1.7 and earlier, NovaSeq Control Software 1.8, iScan Control Software 4.0.0 and 4.0.5, and all versions on the iSeq 100.
CISA says an attacker could take any action at the operating system level and change settings, configurations, software or data on the instrument through a connected network. Some of those instruments are diagnostic, so the data at stake can include clinical results.
Illumina points users to its UCS Vulnerability Instructions Guide, matched to each system's configuration, and asks them to read it before downloading any software.
A file copy helper is the kind of service nobody inventories. It still listens on the network, and running it with unnecessary privileges turned a convenience into the instrument's largest attack surface.