CISA advisory ICSMA-18-086-01, dated March 27, 2018, covers two vulnerabilities in the Philips Alice 6 System, Version R8.0.3 and prior. Philips reported them itself. Each scores 5.3 under CVSS v3 with a network vector, no privileges and no user interaction.
CVE-2018-5451 (CWE-287) is improper authentication: the software does not prove, or insufficiently proves, that an actor is who it claims to be. CVE-2018-7498 (CWE-311) is missing encryption of sensitive data. CISA's risk evaluation says exploitation may let an attacker see usernames, passwords and personal data.
The two flaws compound. Without node authentication, a malicious machine can stand in for a trusted one, and without encryption the data it receives arrives readable. CISA lists disclosure of personal data by replacing a trusted node with a malicious node as the result, along with data that can be altered or corrupted in transit.
Philips said it would notify users, schedule updates and bring every Alice 6 at R8.0.3 or prior to R8.0.4, with the new version and documentation planned for Q1 of 2019. Users with installation questions go to their local Philips service support team.
A clinical system that trusts any machine claiming to be a peer is only as safe as the network it sits on. Segment it until R8.0.4 is installed.