PROService
Remote service and monitoring tool. Attributes come from Beckman Coulter's own public documentation for this product.
Attributes
Signals
Properties (MITRE EMB3D)
PID-27 Device includes support for firmware/software updates PID-275 Device includes remotely-initiated firmware/software updates PID-33 Device includes interactive applications, services, or user interfaces PID-41 Device exposes remote network services PID-411 Device exposes remote services with the ability to send, receive, view or modify sensitive information or configurations PID-4113 Device includes cryptographic functions for sensitive data, such as encryption or authentication PID-42 Device includes procedure to forward or route network messages
Platform
Web server: Microsoft IIS
Attributes: materials
IIS
Attributes: component types
API Firmware Generic Application Generic Hardware Library Local Application Local Service User Interface
Attributes: interface types
Cellular Ethernet TCP TLS WiFi
Attributes: assets
Software Configuration Software Update Telemetry Data
Attributes: dataflows
Beckman Coulter servers to instrument (software, database, help file and reagent file/settings updates) Beckman Coulter service staff to instrument (remote setup/configuration) Beckman Coulter service staff to instrument (remote upload of instrument data) Instrument workstation to RMS RAP box (instrument performance data) RAP box to Beckman Coulter servers (encrypted instrument data over HTTPS) RAP box to facility proxy server
Matched threat intel
Top 20 of 47
Advisories in our coverage that share attributes with this product, strongest first: a named product or component, the same device type or lineage, then the most shared attributes. A shared attribute is a lead, not a finding. The status column shows only what Beckman Coulter has said.
| # | Advisory | Date | Shared | Beckman Coulter status |
|---|---|---|---|---|
| 1 | Carestream Vue RIS leaks technical details in HTTP 500 error pages | Oct 4, 2018 | Same named componentIIS (= 7.5)3 shared attributes: Local Application, TCP, IIS | No response from manufacturer |
| 2 | Roche Accu-Chek and CoaguChek point of care handhelds expose a weakly protected service interface | Nov 6, 2018 | Same property (MITRE EMB3D)PID-2710 shared attributes: Wired network, Wireless connectivity, Field-updatable firmware, Device exposes remote network services (PID-41) +6 | No response from manufacturer |
| 3 | Philips IntelliVue MX40 WLAN monitors can drop local alarming after 802.11 faults | Sep 12, 2017 | Same signals5 shared attributes: Patient monitoring family, Monitor, Wireless connectivity, Firmware +1 | No response from manufacturer |
| 4 | Canon Medical flags FragAttacks Wi-Fi flaws in wireless FPD and ultrasound wireless options | Jul 4, 2022 | Same signals2 shared attributes: Wireless connectivity, WiFi | No response from manufacturer |
| 5 | St. Jude Medical Merlin@home transmitters do not verify the Merlin.net endpoint | Jan 9, 2017 | 11 shared attributesRemote, Monitor, Cloud connectivity, Remote access +7 | No response from manufacturer |
| 6 | A patient monitor with a hidden backdoor that phones home | Jan 30, 2025 | 12 shared attributesPatient monitoring family, Monitor, Cloud connectivity, Wireless connectivity +8 | No response from manufacturer |
| 7 | BridgeLink 26.9.0 closes the Mirth Connect flaws CISA flagged, plus eight inherited CVEs | Oct 1, 2026 | 10 shared attributesDevice exposes remote network services (PID-41), Device exposes remote services with the ability to send, receive, view or modify sensitive information or configurations (PID-411), Device includes procedure to forward or route network messages (PID-42), Device includes support for firmware/software updates (PID-27) +6 | No response from manufacturer |
| 8 | A DICOM viewer that did not check its update server certificate | Feb 20, 2025 | 10 shared attributesField-updatable firmware, Cloud connectivity, Wireless connectivity, Device includes support for firmware/software updates (PID-27) +6 | No response from manufacturer |
| 9 | Smiths Medical Medfusion 4000 syringe pump ships hard-coded wireless, FTP and Telnet credentials | Sep 7, 2017 | 9 shared attributesWireless connectivity, Wired network, Firmware, Local Service +5 | No response from manufacturer |
| 10 | Citrix patches two critical NetScaler zero days after Dutch hospitals shut systems down | Sep 27, 2026 | 9 shared attributesRemote, Device exposes remote network services (PID-41), Device includes support for firmware/software updates (PID-27), Device includes procedure to forward or route network messages (PID-42) +5 | No response from manufacturer |
| 11 | GE CARESCAPE, ApexPro and CIC monitoring systems carry hard-coded SMB credentials and five more critical flaws | Jan 23, 2020 | 9 shared attributesPatient monitoring family, Monitor, Wired network, Remote access +5 | No response from manufacturer |
| 12 | Philips IntelliVue WLAN modules accept malicious firmware through a hard-coded FTP password | Sep 12, 2019 | 7 shared attributesPatient monitoring family, Monitor, Wireless connectivity, Field-updatable firmware +3 | No response from manufacturer |
| 13 | Baxter warns of a Newmar radio PIM file flaw in Connex vital signs monitors | Feb 18, 2026 | 7 shared attributesPatient monitoring family, Monitor, Wireless connectivity, Field-updatable firmware +3 | No response from manufacturer |
| 14 | Dräger Infinity Delta monitors allow kiosk escape, reboot loops and unauthenticated log access | Jan 22, 2019 | 7 shared attributesPatient monitoring family, Monitor, Wired network, User Interface +3 | No response from manufacturer |
| 15 | Baxter Welch Allyn Connex Spot Monitor used a default cryptographic key | May 30, 2024 | 6 shared attributesPatient monitoring family, Monitor, Field-updatable firmware, Wireless connectivity +2 | No response from manufacturer |
| 16 | bioMérieux says three Access:7 flaws reach VILINK with Axeda, rated low risk behind the firewall | Mar 8, 2022 | 7 shared attributesRemote, Service, Remote access, Device includes remotely-initiated firmware/software updates (PID-275) +3 | No response from manufacturer |
| 17 | Philips lists Zenition 70 and VitalSky under Terrapin SSH flaw CVE-2023-48795 | Apr 8, 2024 | 6 shared attributesPatient monitoring family, Remote, Monitor, Remote access +2 | No response from manufacturer |
| 18 | Philips lists MR, SPECT, ultrasound and telemetry products in Urgent/11 VxWorks advisory | Aug 1, 2019 | 6 shared attributesPatient monitoring family, Wired network, Device exposes remote network services (PID-41), Library +2 | No response from manufacturer |
| 19 | bioMérieux finds Ripple20 flaws in networked HP printers shipped with its systems | Oct 14, 2020 | 6 shared attributesWired network, Field-updatable firmware, Device includes support for firmware/software updates (PID-27), Firmware +2 | No response from manufacturer |
| 20 | Carestream DRX wireless detectors affected by KRACK WPA2 key reinstallation flaws | Nov 1, 2017 | 6 shared attributesWireless connectivity, Device includes cryptographic functions for sensitive data, such as encryption or authentication (PID-4113), Generic Hardware, Firmware +2 | No response from manufacturer |
Threats and weaknesses in matched advisories
MITRE EMB3D threats
TID-115 Firmware/Data Extraction via Hardware Interface TID-116 Latent Privileged Access Port TID-119 Latent Hardware Debug Port Allows Memory/Code Manipulation TID-202 Exploitable System Network Stack Component TID-204 Untrusted Programs Can Access Privileged OS Functions TID-205 Existing OS Tools Maliciously Used for Device Manipulation TID-210 Device Vulnerabilities Unpatchable TID-211 Device Allows Unauthenticated Firmware Installation TID-213 Faulty FW/SW Update Integrity Verification TID-215 Unencrypted SW/FW Updates TID-219 OS/Kernel Privilege Escalation TID-221 Authentication Bypass By Message Replay TID-222 Critical System Service May Be Disabled TID-224 Excessive Access via Software Diagnostic Features TID-226 Device leaks security information in logs TID-304 Manipulate Runtime Environment TID-310 Remotely Accessible Unauthenticated Services TID-311 Default Credentials TID-313 Unauthenticated Session Changes Credential TID-316 Incorrect Certificate Verification Allows Authentication Bypass TID-317 Predictable Cryptographic Key TID-320 SQL Injection TID-324 HTTP Direct Object Reference TID-326 Insecure Deserialization TID-327 Out of Bounds Memory Access TID-328 Hardcoded Credentials TID-329 Improper Password Storage TID-401 Undocumented Protocol Features TID-404 Remotely Triggerable Deadlock/DoS TID-405 Network Stack Resource Exhaustion TID-406 Unauthorized Messages or Connections TID-408 Unencrypted Sensitive Data Communication TID-411 Weak/Insecure Cryptographic Protocol
CWE weaknesses
CWE-1104 Use of Unmaintained Third Party Components CWE-1188 Insecure Default Initialization of Resource CWE-1191 On-Chip Debug and Test Interface With Improper Access Control CWE-122 Heap-based Buffer Overflow CWE-125 Out-of-bounds Read CWE-1313 Hardware Allows Activation of Test or Debug Logic at Runtime CWE-1329 Reliance on Component That is Not Updateable CWE-1390 Weak Authentication CWE-1392 Use of Default Credentials CWE-1395 Dependency on Vulnerable Third-Party Component CWE-20 Improper Input Validation CWE-200 Exposure of Sensitive Information to an Unauthorized Actor CWE-276 Incorrect Default Permissions CWE-284 Improper Access Control CWE-287 Improper Authentication CWE-294 Authentication Bypass by Capture-replay CWE-306 Missing Authentication for Critical Function CWE-319 Cleartext Transmission of Sensitive Information CWE-321 Use of Hard-coded Cryptographic Key CWE-326 Inadequate Encryption Strength CWE-345 Insufficient Verification of Data Authenticity CWE-400 Uncontrolled Resource Consumption CWE-427 Uncontrolled Search Path Element CWE-476 NULL Pointer Dereference CWE-494 Download of Code Without Integrity Check CWE-506 Embedded Malicious Code CWE-522 Insufficiently Protected Credentials CWE-639 Authorization Bypass Through User-Controlled Key CWE-787 Out-of-bounds Write CWE-798 Use of Hard-coded Credentials CWE-912 Hidden Functionality CWE-916 Use of Password Hash With Insufficient Computational Effort