Philips published an advisory on August 1, 2019 on Urgent/11, the set of 11 vulnerabilities Armis disclosed in Wind River's VxWorks real-time operating system and its IPnet TCP/IP stack. Philips says the flaws could lead to remote code execution and allow an attacker to take over a whole system without interacting with the user.

The Philips list covers Achieva and Achieva 3.0T (R5.3, R5.4 and higher), Ingenia (R4, R5.3, R5.4 and higher), Multiva and Prodiva (R5.3, R5.4), BrightView SPECT 1.x, BrightView X and XCT 2.x, HDI 3000 and HDI 3500, Zenition, IntelliSpace Breast, the GEOPC component of Allura and Azurion, and the Smart-hopping Access Point Controller for MX40 and telemetry products.

A network stack bug sits below every application control, so authentication on the device does not help. The telemetry access point controller is the entry to watch: it carries patient alarms.