CISA published advisory ICSMA-19-255-01 on September 12, 2019, covering two vulnerabilities in the WLAN modules of Philips IntelliVue MP and MX patient monitors. Affected are WLAN Version A with firmware A.03.09 (MP20 to MP90, MP5 and MP5SC) and WLAN Version B with firmware A.01.09 (MP2/X2 and MX800, MX700, MX600). Shawn Loveric of Finite State reported them.

CVE-2019-13530 (CWE-259, CVSS v3 6.4) is a hard-coded password: an attacker can use it to log in over FTP and upload malicious firmware. CVE-2019-13534 (CWE-494, CVSS 6.4) closes the loop, since the module runs downloaded code without verifying its origin or integrity.

The impact Philips described is corrupted WLAN firmware and disrupted data flow over the wireless module, producing an inoperative condition alert at the monitor and the Central Station. The vector needs adjacent network access and high privileges, which in practice means the wireless network the monitors join.

WLAN Version B is obsolete, and Philips recommends replacing it with the Version C module, whose current firmware B.00.31 is not vulnerable. For Version A, Philips estimated a software patch on InCenter by the end of 2019. Meanwhile, wireless access should require authentication such as WPA2, and a firewall rule should restrict FTP on the wireless network.

A hard-coded password plus unsigned firmware is two mistakes, and fixing either one breaks the attack.