CISA published advisory ICSMA-17-255-01 on September 12, 2017, covering the Philips IntelliVue MX40 Patient Worn Monitor, WLAN models only, on software earlier than B.06.18. It lists two vulnerabilities, each with a CVSS v3 base score of 6.5: CVE-2017-9657 (CWE-460, improper cleanup on thrown exception) and CVE-2017-9658 (CWE-755, improper handling of exceptional conditions).

CVE-2017-9657 is a state mismatch. Under specific 802.11 conditions the MX40 can partially re-associate with the central monitoring station. The central station then shows the monitor as disconnected and expects it to fall back to local monitoring with audio and screen on, while the MX40 stays in telemetry mode with audio and screen off.

The clinical consequence is the one the advisory spells out: if the patient has an alarm event and staff expect the MX40 to alarm locally, treatment can be delayed. Nobody has to compromise anything for this to happen. A degraded wireless environment is enough.

CVE-2017-9658 concerns certain 802.11 management messages that trip the access point blacklisting defenses when they should not. The monitor then needs hospital staff to reset it and re-establish its Wi-Fi connection, unless it finds another access point in range. Until reset, it can remain in local monitoring mode.

Philips told CISA it planned an additional MX40 software update in 2017 for the exception handling issue. A telemetry device has to agree with the central station about who is alarming; any state where both sides assume the other one is responsible is a patient safety defect.