CISA published advisory ICSMA-17-009-01A on January 9, 2017, covering one vulnerability in St. Jude Medical's Merlin@home transmitter. CVE-2017-5149 (CWE-300, Channel Accessible by Non-Endpoint) carries a CVSS v3 base score of 8.9. Affected are the inductive EX1100 models, the EX1100 with MerlinOnDemand capability and the RF EX1150 models, all on transmitter software below version 8.2.2.
The flaw sits in the link between the transmitter and St. Jude Medical's web site, Merlin.net. The identities of the endpoints on that channel are not verified, so a remote attacker who gets onto the path could read or influence the traffic. The vector rates attack complexity high and the integrity and availability impact high, with a scope change.
Version 8.2.2 fixes it, and nobody installs it by hand. St. Jude Medical rolls the update out automatically over several months whenever a transmitter is connected to an Ethernet, WiFi, cellular network or a landline, and recommends keeping the units powered and connected at all times.
A transmitter unplugged in a drawer never receives the fix, so the patch rate depends on patients as much as on the manufacturer. Mutual authentication between a device and its back end belongs in the first release, because retrofitting it costs a months-long rollout.