CISA published advisory ICSMA-26-169-01 on June 18, 2026, covering the Apollo Pharmacy Blood Glucose Monitoring System, Model No. APG-01 BT, firmware 0x0110_v1.1.0. Rishitha Pucchakayala of the Centre for Development of Advanced Computing (Hyderabad) reported two vulnerabilities, each scored 6.5 under CVSS v3.1.

CVE-2026-50034 (CWE-319) is cleartext transmission. An attacker within BLE range can passively intercept the traffic and read sensitive health information, including glucose measurement values. CVE-2026-52866 (CWE-862) is missing authorization on the connection itself: an attacker can occupy the meter's only BLE connection slot and lock out the legitimate app.

Both need radio proximity and neither is remotely exploitable, per CISA. The second matters more than its confidentiality-free score suggests, since a meter that cannot sync leaves the patient's app without readings.

Apollo Pharmacy did not respond to CISA's requests to coordinate, so there is no fix. CISA directs users to contact Apollo Pharmacy and to its Understanding Bluetooth Technology guidance. A consumer meter that pairs with anything nearby and talks in the clear has no security boundary at all.