CISA published advisory ICSMA-17-229-01 on August 17, 2017, covering two vulnerabilities in the Philips DoseWise Portal (DWP), a web-based application. The affected versions are 1.1.7.333 and 2.1.1.3069.
CVE-2017-9656 (CWE-798) scores 9.1 under CVSS v3. The backend database uses hard-coded credentials for a database account with privileges over the confidentiality, integrity and availability of that database. Those credentials sit in backend system files.
CVE-2017-9654 (CWE-312, CVSS 6.5) is the companion flaw: the application stores login credentials in clear text within the same kind of backend system files. An attacker who reaches those files collects both the user logins and the database account.
The critical score carries a precondition. Elevated privileges are first required to reach the backend system files, which is why the vector reads PR:H. Once that step is done, a remote attacker gains access to the DWP database, and that database contains PHI.
Philips will support every user of version 1.1.7.333 to reconfigure the installation, changing and fully encrypting all stored passwords. The advisory text names no separate step for 2.1.1.3069.
A database password baked into the product is the same on every install, so one leaked config file opens every site. Generate the secret per installation and keep it out of readable files.