CISA published advisory ICSMA-19-190-01 on July 9, 2019, for CVE-2019-10966 in GE Healthcare anesthesia devices. The listed products are the Aespire 7100, 7900, 100 and Protiva Carestation View; the Aestiva 7100 and 7900 MRI; the Aisys CS2, Avance, Amingo and Avance CS2; and the Carestation 620, 650 and 650c. Elad Luz of CyberMDX reported it.

The weakness is improper authentication (CWE-287) with a CVSS v3 base score of 5.3. It exists when the devices' serial ports are connected through an added, unsecured terminal server to a TCP/IP network. An attacker who reaches that terminal server could remotely modify device configuration and silence alarms.

CISA's risk evaluation ties the exposure to the configuration of certain terminal server implementations that extend the anesthesia device serial ports to TCP/IP networks. The fix therefore sits largely with whoever installed and configured the terminal server.

GE Healthcare recommends secure terminal servers, which provide encryption, VPN, user authentication, network controls, logging, audit capability and secure configuration management. It also recommends governance and secure deployment measures such as network segmentation, VLANs and device isolation, and said it would post updates for affected users on its security page.

A flaw rated 5.3 that can silence alarms on an anesthesia delivery system deserves priority well above its score.