Similar Products
A security flaw in the server behind GE Revolution CT scanners. Products whose own public documentation makes them the same type of product as the one this advisory names, gives them the attack surface the flaw needs, or both. Within each group a named product or component and the same device type or lineage come first, then the most shared attributes, rarer ones counting for more.
This advisory's attributes
Ranked products
Two reasons a product is similar. The same type of product: a peer of the product this advisory names, whatever technology it uses. The same attack surface: any product whose documentation states what this flaw needs. A product with both comes first. Other shared attributes lists products that overlap on attributes but have not been scored on either reason. A shared attribute is a lead, not a finding. The status column shows only what each manufacturer has said.
| # | Product | Match | Shared attributes | Manufacturer status |
|---|---|---|---|---|
| 1 | SCENARIA View Phase 5.0Fujifilm Healthcare, System, X-Ray, Tomography, Computed | Same FDA device type6 shared attributes | Device typeimaging family SignalRemote access Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| # | Product | Match | Shared attributes | Manufacturer status |
|---|---|---|---|---|
| 1 | LOGIQ eGE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 11 shared attributes | Device typeimaging family imag SignalCloud connectivity Field-updatable firmware Hospital network Handles patient data Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) ComponentCloud Service AssetPatient Data | No response from manufacturer |
| 2 | ACUSON Sequoia Diagnostic Ultrasound SystemSiemens Healthineers, System, Imaging, Pulsed Doppler, Ultrasonic | 11 shared attributes | Device typeimaging family imag SignalRemote access Field-updatable firmware Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 3 | Versana PremierGE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 10 shared attributes | Device typeimaging family imag SignalCloud connectivity Remote access Hospital network Handles patient data Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 4 | Vscan AirGE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 10 shared attributes | Device typeimaging family imag SignalCloud connectivity Remote access Field-updatable firmware Hospital network Handles patient data PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 5 | Vivid iqGE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 10 shared attributes | Device typeimaging family imag SignalField-updatable firmware Hospital network Handles patient data Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) InterfaceUnknown AssetPatient Data | No response from manufacturer |
| 6 | NovoPathAI, Digital Pathology Image Viewing And Management Software | 9 shared attributes | Device typeimaging family SignalCloud connectivity Off-the-shelf software Handles patient data Patient or study database PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) ComponentCloud Service AssetPatient Data | No response from manufacturer |
| 7 | Vivid S60N, Vivid S70NGE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 10 shared attributes | Device typeimaging family imag SignalField-updatable firmware Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 8 | Voluson Expert 18; Voluson Expert 20; Voluson Expert 22GE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 9 shared attributes | Device typeimaging family imag SignalOff-the-shelf software Hospital network Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) InterfaceUnknown AssetPatient Data | No response from manufacturer |
| 9 | Hepatic VCARGE HealthCare, System, X-Ray, Tomography, Computed | 8 shared attributes | Device typeimaging family SignalHospital network Patient or study database PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) InterfaceUnknown AssetPatient Data MaterialAW Server | No response from manufacturer |
| 10 | Harmony Referral SystemTopcon Healthcare, System, Image Management, Ophthalmic | 9 shared attributes | Device typeimaging family SignalRemote access Off-the-shelf software Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) | No response from manufacturer |
| 11 | Infinity Gateway SuiteDräger, System, Network And Communication, Physiological Monitors | 10 shared attributes | SignalRemote access Off-the-shelf software Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) InterfaceUnknown AssetPatient Data | No response from manufacturer |
| 12 | Aperio GT 450 DX; Aperio GT 180 DXLeica Biosystems, Whole Slide Imaging System | 9 shared attributes | Device typeimaging family imag SignalRemote access Off-the-shelf software Hospital network Handles patient data PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 13 | Tandem Mobi insulin pump with interoperable technologyTandem Diabetes Care, Alternate Controller Enabled Insulin Infusion Pump | 9 shared attributes | SignalCloud connectivity Remote access Field-updatable firmware Handles patient data PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) ComponentCloud Service InterfaceUnknown AssetPatient Data | No response from manufacturer |
| 14 | Assert-IQ Insertable Cardiac MonitorAbbott, Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection) | 9 shared attributes | SignalCloud connectivity Remote access Field-updatable firmware Handles patient data Patient or study database PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) ComponentCloud Service AssetPatient Data | No response from manufacturer |
| 15 | NAEOTOM AlphaSiemens Healthineers, System, X-Ray, Tomography, Computed | 9 shared attributes | Device typeimaging family SignalRemote access Off-the-shelf software Field-updatable firmware Hospital network Handles patient data PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 16 | ViewPoint 6GE HealthCare, System, Image Processing, Radiological | 9 shared attributes | Device typeimaging family SignalOff-the-shelf software Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 17 | Synapse PACSFujifilm Healthcare, Automated Radiological Image Processing Software | 9 shared attributes | Device typeimaging family SignalOff-the-shelf software Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 18 | BodyGuardian Remote Monitoring SystemBoston Scientific, Detector And Alarm, Arrhythmia | 9 shared attributes | SignalCloud connectivity Remote access Off-the-shelf software Handles patient data Patient or study database PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) InterfaceUnknown AssetPatient Data | No response from manufacturer |
| 19 | VesselIQ XpressGE HealthCare, System, X-Ray, Tomography, Computed | 7 shared attributes | Device typeimaging family SignalOff-the-shelf software Hospital network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data MaterialAW Server | No response from manufacturer |
| 20 | Philips IntelliSpace CardiovascularPhilips, System, Image Processing, Radiological | 9 shared attributes | SignalCloud connectivity Remote access Hospital network Handles patient data Patient or study database PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) ComponentCloud Service AssetPatient Data | No response from manufacturer |
| 21 | aPROMISE XExini Diagnostics, System, Image Processing, Radiological | 9 shared attributes | Device typeimaging family SignalCloud connectivity Hospital network Handles patient data Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 22 | Vivid T8, Vivid T9GE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 9 shared attributes | Device typeimaging family imag SignalField-updatable firmware Hospital network Handles patient data Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 23 | da Vinci SP Surgical SystemIntuitive, System, Surgical, Computer Controlled Instrument | 8 shared attributes | SignalCloud connectivity Remote access Field-updatable firmware Patient or study database Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) InterfaceUnknown | No response from manufacturer |
| 24 | Aperio GT 450 DXLeica Biosystems, Whole Slide Imaging System | 8 shared attributes | Device typeimaging family imag SignalRemote access Off-the-shelf software Hospital network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |
| 25 | Venue FitGE HealthCare, System, Imaging, Pulsed Doppler, Ultrasonic | 8 shared attributes | Device typeimaging family imag SignalOff-the-shelf software Hospital network Wired network PropertyDevice exposes remote network services (PID-41) Application-level software is present and running on the device (PID-31) AssetPatient Data | No response from manufacturer |