CISA published advisory ICSMA-17-292-01 on October 19, 2017, covering the Boston Scientific ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model 3120. All versions are affected by two vulnerabilities, each scored 4.6 under CVSS v3.

CVE-2017-14014 (CWE-321) is a hard-coded cryptographic key. The PRM uses it to encrypt PHI before writing it to removable media, so anyone who recovers the key from one unit can decrypt exports from every unit. CVE-2017-14012 (CWE-311) is simpler: the device does not encrypt PHI at rest at all.

Both CVSS vectors start with physical access. An attacker has to hold the programmer or the media it wrote, which keeps the scores low and the exposure tied to loss, theft and disposal rather than network attack. A programmer that leaves a clinic for repair or resale carries its patient records with it.

Boston Scientific said it will not issue a product update for the Model 3120 and pointed users to compensating controls. Physical custody of the programmer and its media is the only control left, so treat both like a paper chart.