CISA published advisory ICSMA-20-170-02 on June 18, 2020, later revised as Update B, covering three vulnerabilities in Baxter PrismaFlex (all versions) and PrisMax (versions before 3.x). Baxter reported the issues itself.

CVE-2020-12036 (CWE-319, CVSS v3 6.5) concerns the data path to hospital systems. The devices do not implement data-in-transit encryption such as TLS when configured to send treatment data to a Patient Data Management System (PDMS) or an Electronic Medical Record (EMR) system, so an attacker on the network can observe that data.

CVE-2020-12035 (CWE-287, CVSS v3 7.6) carries the higher score. The same connection requires no authentication, which could let an attacker modify treatment status information. CVE-2020-12037 (CWE-259, CVSS v3 5.4) is a hard-coded service password on PrismaFlex that opens biomedical information, device settings, calibration settings and network configuration to anyone with physical access, who could then change settings and calibration.

Baxter's own recommendations add a clinical angle. A PDMS or EMR system that is not compatible with PrismaFlex and PrisMax can present erroneous data, so the integration has to be verified and its risks analyzed, and later changes to the enterprise network can introduce new risks.

For PrisMax, Baxter recommends upgrading to PrisMax v3 with the Digital Communication Module (DCM), which supports a mutually authenticated TLS tunnel to a PDMS or EMR system capable of TLS 1.2. PrismaFlex software 8.2x adds the option to set a device-specific service password, though 8.2x is not available in all regions, including the United States.

Unauthenticated writes into a clinical record are an integrity problem before they are a privacy problem.