DxS IntelliServe
Remote service and diagnostics platform. Attributes come from Beckman Coulter's own public documentation for this product.
Attributes
Properties (MITRE EMB3D)
PID-27 Device includes support for firmware/software updates PID-275 Device includes remotely-initiated firmware/software updates PID-41 Device exposes remote network services PID-411 Device exposes remote services with the ability to send, receive, view or modify sensitive information or configurations
Attributes: component types
API Cloud Service Firmware Local Service
Attributes: assets
Software Update Telemetry Data
Attributes: dataflows
DxS IntelliServe to Analyzer (automatic software, security and test menu updates) Instrument to DxS IntelliServe (real-time data gathering and file transfers) Instrument to DxS IntelliServe cloud (critical device information, real-time monitoring data)
Matched threat intel
11
Advisories in our coverage that share attributes with this product, strongest first: a named product or component, the same device type or lineage, then the most shared attributes. A shared attribute is a lead, not a finding. The status column shows only what Beckman Coulter has said.
| # | Advisory | Date | Shared | Beckman Coulter status |
|---|---|---|---|---|
| 1 | A DICOM viewer that did not check its update server certificate | Feb 20, 2025 | Same property (MITRE EMB3D)PID-2756 shared attributes: Field-updatable firmware, Cloud connectivity, Device includes support for firmware/software updates (PID-27), Device includes remotely-initiated firmware/software updates (PID-275) +2 | No response from manufacturer |
| 2 | Medtronic 2090 CareLink and 29901 Encore programmers accept updates without verifying their origin | Feb 27, 2018 | Same property (MITRE EMB3D)PID-2755 shared attributes: Field-updatable firmware, Cloud connectivity, Device includes support for firmware/software updates (PID-27), Cloud Service +1 | No response from manufacturer |
| 3 | A hard-coded VNC password on a bioreactor, base score 9.8 | May 26, 2026 | Same property (MITRE EMB3D)PID-4116 shared attributes: Remote access, Field-updatable firmware, Device exposes remote network services (PID-41), Device exposes remote services with the ability to send, receive, view or modify sensitive information or configurations (PID-411) +2 | No response from manufacturer |
| 4 | MicroDicom DICOM Viewer 2024.03 fails to validate its update server certificate | Feb 6, 2025 | Same property (MITRE EMB3D)PID-2754 shared attributes: Field-updatable firmware, Device includes support for firmware/software updates (PID-27), Cloud Service, Software Update | No response from manufacturer |
| 5 | St. Jude Medical Merlin@home transmitters do not verify the Merlin.net endpoint | Jan 9, 2017 | 8 shared attributesRemote, Cloud connectivity, Remote access, Field-updatable firmware +4 | No response from manufacturer |
| 6 | bioMérieux says three Access:7 flaws reach VILINK with Axeda, rated low risk behind the firewall | Mar 8, 2022 | 6 shared attributesRemote, Service, Remote access, Device includes remotely-initiated firmware/software updates (PID-275) +2 | No response from manufacturer |
| 7 | BD rates Meltdown and Spectre low impact across its instrument and Pyxis lines | Jan 5, 2018 | 4 shared attributesLab and diagnostics family, Diagnostic, Firmware, Software Update | No response from manufacturer |
| 8 | Carestream patches Axeda clients behind Smart Link remote management for Access:7 | Mar 8, 2022 | 5 shared attributesRemote, Service, Remote access, Local Service +1 | No response from manufacturer |
| 9 | Cepheid GeneXpert Xpertise 7.1 needs .NET October 2025 update for CVE-2025-55248 | Mar 11, 2026 | 3 shared attributesLab and diagnostics family, Diagnostic, Software Update | No response from manufacturer |
| 10 | Hologic removes PTC Axeda agent from systems running Hologic Connect | Mar 15, 2022 | 4 shared attributesRemote, Service, Remote access, Local Service | No response from manufacturer |
| 11 | Philips e-Alert units through R2.1 send communication unencrypted | Aug 30, 2018 | 3 shared attributesRemote, Service, Remote access | No response from manufacturer |
Threats and weaknesses in matched advisories
MITRE EMB3D threats
TID-103 Microarchitectural Side Channels TID-202 Exploitable System Network Stack Component TID-210 Device Vulnerabilities Unpatchable TID-211 Device Allows Unauthenticated Firmware Installation TID-213 Faulty FW/SW Update Integrity Verification TID-215 Unencrypted SW/FW Updates TID-222 Critical System Service May Be Disabled TID-310 Remotely Accessible Unauthenticated Services TID-311 Default Credentials TID-316 Incorrect Certificate Verification Allows Authentication Bypass TID-323 Path Traversal TID-328 Hardcoded Credentials TID-329 Improper Password Storage TID-404 Remotely Triggerable Deadlock/DoS TID-406 Unauthorized Messages or Connections TID-408 Unencrypted Sensitive Data Communication TID-411 Weak/Insecure Cryptographic Protocol
CWE weaknesses
CWE-1104 Use of Unmaintained Third Party Components CWE-1329 Reliance on Component That is Not Updateable CWE-1392 Use of Default Credentials CWE-1395 Dependency on Vulnerable Third-Party Component CWE-319 Cleartext Transmission of Sensitive Information CWE-326 Inadequate Encryption Strength CWE-345 Insufficient Verification of Data Authenticity CWE-494 Download of Code Without Integrity Check CWE-522 Insufficiently Protected Credentials CWE-798 Use of Hard-coded Credentials