CISA published advisory ICSMA-21-322-02 on November 18, 2021, covering three vulnerabilities in the Philips Patient Information Center iX (PIC iX) central station and the Efficia CM Series patient monitors. Younes Dragoni, Andrea Palanca and Ivan Speziale of Nozomi Networks reported them.
CVE-2021-43548 (CWE-20, CVSS v3 6.5) is improper input validation in PIC iX C.02 and C.03, reachable from an adjacent network with no credentials. Its impact is availability: a temporary interruption of viewing physiological data at the central station.
CVE-2021-43552 (CWE-321, CVSS 6.1) is a hard-coded cryptographic key in PIC iX B.02, C.02 and C.03, which makes encrypted data easier to recover. CVE-2021-43550 (CWE-327, CVSS 5.9) is a broken or risky algorithm protecting traffic between PIC iX C.02 and C.03 and Efficia CM Series revisions A.01 to C.0x and 4.0.
Philips said exploitation does not enable modification of point-of-care devices. The exposure is patient data and the central view of it. A key compiled into every installation is a shared secret across every hospital that runs the product, and one extraction serves them all.
The input validation fix shipped in Q3 2021 as PIC iX C.03.06. Philips planned to remediate the two cryptographic flaws by the end of Q2 2023, a long tail that leaves the interim controls doing the work.
Those controls: keep the patient monitoring network physically or logically isolated from the hospital LAN, filter it to necessary ports and addresses, leave BitLocker Drive Encryption on, sanitize media to NIST SP 800-88 before disposal, and lock down any archive exported with patient information.