CISA published advisory ICSMA-18-233-01 on August 21, 2018, covering the Philips IntelliVue Information Center iX (PIIC iX) version B.02. The single vulnerability, CVE-1999-0103 (CWE-400), scores 5.7 under CVSS v3. An anonymous party and Philips are credited.
An attacker can exhaust the system by sending multiple initial UDP requests. CISA's risk evaluation says the operating system becomes unresponsive under the attack, which affects the application's ability to meet its intended use. The vector requires adjacent network access and low privileges.
PIIC iX is a central monitoring station, so the clinical cost of an outage is the loss of the central view of patient monitors, not a stopped device at the bedside. The attacker has to be on the same network segment, which is exactly where a compromised workstation on a flat clinical VLAN already sits.
Philips put mitigations in place and pointed users to the device labeling, instructions for use and service guides for compensating controls. It committed to a patch for all PIIC iX B.02 users by the end of November 2018.
A CVE number from 1999 on a 2018 monitoring product says the network stack inherited old behavior nobody switched off. Close the UDP services the station does not use.