CISA published advisory ICSMA-23-180-01 on June 29, 2023, covering CVE-2023-31222 in the Medtronic Paceart Optima system, versions 1.11 and prior. The flaw is deserialization of untrusted data (CWE-502), scored 9.8 under CVSS v3. Medtronic reported it.
Exposure depends on configuration. If a healthcare delivery organization has enabled the optional Paceart Messaging Service, an unauthorized user can send specially crafted messages to the system and get remote code execution, a denial of service, or both.
Code execution on Paceart Optima could mean deletion, theft or modification of the system's cardiac device data, or using the server as a base for further network penetration. A denial of service could slow the system or leave it unresponsive. CISA reported no known public exploits.
Medtronic's fix is version 1.12, scheduled through Medtronic. Until then, sites running a combined Application and Integration Server should call Paceart Optima technical support for immediate steps.
For other configurations the mitigation is to switch the feature off: stop the Paceart Messaging Service on the Application Server and set its startup type to Disabled, then remove the Message Queuing feature through Server Manager. The vulnerability stays mitigated as long as the service remains disabled.
An optional service that nobody uses should never be listening. Ship it disabled and make enabling it a deliberate act.