CISA published advisory ICSMA-19-080-01 on March 21, 2019, later revised through Update C, covering two vulnerabilities in the Medtronic Conexus radio frequency telemetry protocol. Conexus carries data between Medtronic implanted cardiac devices and the CareLink 2090 Programmer, the CareLink Monitor 2490C and the MyCareLink Monitor 24950 and 24952.

All versions of the affected implants are listed. On the CRT-D side: Amplia, Brava, Claria, Compia, Concerto, Concerto II, Consulta and Viva. On the ICD side: Evera, Mirro, Mirro MRI, Nayamed ND, Primo, Secura, Virtuoso, Virtuoso II and Visia AF. Maximo II and Protecta appear in both ICD and CRT-D form. Credited researchers include Peter Morgan of Clever Security, Dave Singelée and Bart Preneel of KU Leuven, Eduard Marin and Tom Chothia of the University of Birmingham, Flavio D. Garcia, and Rik Willems of University Hospital Gasthuisberg Leuven.

CVE-2019-6538 (CWE-284, CVSS v3 9.3) is the core finding: Conexus implements neither authentication nor authorization. An attacker with adjacent short-range access, while the product's radio is on, can inject, replay, modify or intercept telemetry data. The protocol can read and write memory values on the implant, so that access reaches the implanted device's memory directly.

CVE-2019-6540 (CWE-319, CVSS v3 6.5) is the companion flaw. Conexus does not implement encryption, so an attacker in range can listen to communications, including sensitive data.

Exploitation has real preconditions. The attacker needs an RF device able to speak Conexus, such as a monitor, programmer or software-defined radio, plus short-range proximity and a product whose radio is active. Before implant and during clinic follow-ups, a session must first be started by an inductive protocol; outside those settings the implant turns its radio on only briefly for scheduled transmissions and notifications. CISA states the flaws are not exploitable remotely.

Medtronic added controls for monitoring and responding to improper use of Conexus by the implants. It developed patches for a subset of models, installed during regular office visits, and said patches for more models would follow through future updates, assuming regulatory approval.

An unauthenticated protocol with memory write access to a defibrillator is a design decision that later patches only partly unwind. Authentication on the telemetry link has to ship in the first release.