CISA published advisory ICSMA-20-049-02 on February 18, 2020, for a restricted desktop escape in GE Healthcare ultrasound products. CVE-2020-6977 (CWE-693, protection mechanism failure) carries a CVSS v3.1 base score of 8.4. Researchers credited in the advisory include Marc Ruef and Rocco Gagliardi of scip AG, Michael Aguilar of Secureworks and Jonathan Bouman of Protozoan.nl.
Six GE ultrasound lines are affected: Invenia ABUS Scan station (not the VScan line), LOGIQ (except LOGIQ 100 Pro), Venue (except Venue 40 R1-3 and Venue 50 R4-5), Versana Essential, Vivid (except EchoPAC) and Voluson (except ImageVault). All versions are listed.
The flaw is in Kiosk Mode, the locked-down shell that keeps sonographers inside the ultrasound application. Specially crafted inputs at the console break out of it and land the user on the underlying operating system. The vector is local, with no privileges or user interaction required, and full confidentiality, integrity and availability impact.
GE's mitigations are physical. Restrict physical access to the devices, and where available turn on the system lock password in the Administration GUI menu, which requires a password before anyone can use the system. Questions go to a GE Healthcare service representative or the GE product security portal.
Ultrasound carts sit in exam rooms and corridors where patients and visitors pass. A kiosk shell is a usability feature that should never be counted as the access control.