CISA published advisory ICSMA-22-251-01 on September 8, 2022, for Baxter Sigma Spectrum 6.x (model 35700BAX) and 8.x (35700BAX2) infusion pumps and Spectrum IQ 9.x (35700BAX3), all through their Wireless Battery Modules (WBM). Deral Heiland of Rapid 7 reported four vulnerabilities.
CVE-2022-26390 (CWE-311, CVSS v3 4.2): the WBM stores network credentials and PHI unencrypted. PHI is only stored in Spectrum IQ pumps using auto programming. Anyone with physical access to a pump or WBM that was not wiped may be able to extract it.
CVE-2022-26394 (CWE-306, CVSS 7.5) is the highest score. The WBM does not perform mutual authentication with the gateway server host, so an attacker can sit in the middle, modify parameters to break the connection, or spoof the server and send crafted data. CVE-2022-26392 (CVSS 3.1) and CVE-2022-26393 (CVSS 5.0) are format string flaws in application messaging that can read WBM memory, and the second can also cause a denial of service.
The physical access case is the one hospitals create themselves. Pumps and battery modules get decommissioned, resold or moved between facilities, and each one can carry the hospital's Wi-Fi credentials out the door.
Baxter says WBM version 20D30 fixes CVE-2022-26393, Spectrum IQ already provides authentication for CVE-2022-26394, software updates to disable Telnet and FTP are in process, and erase instructions are being added to the Spectrum Operator's Manual. It also recommends physical controls, a dedicated VLAN for the infusion system, and the strongest available wireless security such as WPA2 and EAP-TLS.
Treat a pump leaving the building like a laptop leaving the building: wipe it before it goes.