CISA published advisory ICSMA-20-170-04 on June 18, 2020, later revised as Update B, covering six vulnerabilities in Baxter infusion systems: the Sigma Spectrum v6.x (model 35700BAX), Baxter Spectrum v8.x (model 35700BAX2) and v9.x (model 35700BAX3), and the Spectrum Wireless Battery Module (WBM) across firmware from v9 through v22D28. Baxter Healthcare reported them.

On the pump, CVE-2020-12039 (CWE-259, CVSS v3 4.3) is a set of hard-coded passwords that, typed on the keypad, open the biomedical menus: device settings, calibration values and WBM network configuration. CVE-2020-12040 (CWE-319, CVSS 7.3) is an unauthenticated cleartext channel at the application layer for system status and operational data, open to anyone past network defenses for viewing or man-in-the-middle.

The WBM carries the rest. CVE-2020-12045 (CVSS 8.6) is a Telnet service with hard-coded credentials when the module is connected to specific SSIDs. CVE-2020-12041 (CWE-732, CVSS 8.6) means that Telnet interface exposes sensitive data and allows temporary network setting changes and reboots.

CVE-2020-12047 (CVSS 7.3) is an FTP service with hard-coded credentials in the factory-default wireless configuration of WBM v17 onward. CVE-2020-12043 (CWE-672, CVSS 7.3) keeps that FTP service running until the WBM is rebooted.

The advisory lists no fixed version. Baxter's guidance is operational: physical controls around pumps, a dedicated VLAN for the Spectrum system, WPA2 or EAP-TLS wireless security, blocking unexpected traffic such as FTP at the VLAN boundary, and rebooting each WBM after configuration by removing it from the pump for 10 to 15 seconds.

As a last resort, Baxter says the wireless function can be disabled, because the Spectrum was designed to run without network access. The cost is slower drug library updates.

Pulling a battery module for 15 seconds to kill an FTP server is a reasonable fix to put in the biomed checklist, and a strange one to need.