Baxter published a bulletin on July 30, 2020, under ICS advisory ICSA-20-168-01. PrisMax software version 2.1, when connected to an EMR/PDMS via Ethernet, uses GHnet v2, a modified Treck TCP/IP stack, and is affected by 3 of the 19 Ripple20 vulnerabilities: CVE-2020-11907, CVE-2020-11911 and CVE-2020-11914.
Baxter states the impact is limited to network connectivity. Limited still means charting gaps, so check how each unit behaves when its EMR link drops.
Questions & Answers
Is Baxter PrisMax affected by Ripple20?
Baxter says PrisMax software 2.1, when connected to an EMR/PDMS via Ethernet, is affected by three Ripple20 vulnerabilities, CVE-2020-11907, CVE-2020-11911 and CVE-2020-11914, through GHnet v2, a modified Treck TCP/IP stack.
What is the impact of the Ripple20 flaws on Baxter PrisMax 2.1?
Baxter's July 30, 2020 bulletin states the impact of the three Ripple20 vulnerabilities on PrisMax 2.1 is limited to network connectivity.