CISA published advisory ICSMA-17-215-01 on August 3, 2017, covering Siemens molecular imaging products built on Windows XP. The affected lines are the PET/CT Systems, SPECT Systems, SPECT/CT Systems and SPECT Workplaces/Symbia.net, all versions of the Windows XP-based builds.
Two CVEs carry the advisory, and both score 9.8 under CVSS v3. CVE-2008-4250 (CWE-94) lets an unauthenticated remote attacker execute arbitrary code through a crafted remote procedure call request to the Windows server service. CVE-2017-7269 (CWE-119) is a buffer overflow in the WebDAV service: one crafted HTTP request runs code with the permissions of the web server.
Neither flaw is in Siemens code. Both live in the operating system and its bundled web stack, and the advisory record lists them against the SPECT Workplaces/Symbia.net product. The vectors need no credentials and no user interaction, which is the profile worm-style malware goes looking for on flat hospital networks.
CVE-2008-4250 was assigned in 2008, nine years before this advisory. A scanner running in an imaging suite in 2017 that still answered that RPC call tells you how long a medical workstation can sit on an unchanged operating system image.
The CISA record notes that Siemens issued further recommendations to users but does not reproduce a product update in the CSAF data. Any imaging system still on Windows XP belongs on its own segment, with RPC and WebDAV unreachable from anywhere it does not need them.