Philips posted a notice on August 30, 2024 on CVE-2024-38063, a critical TCP/IP remote code execution vulnerability in Microsoft Windows. It affects every system running IPv6, which is enabled by default.
The product list is the longest in this batch of notices. It includes Azurion R2 and R3, Allura 8.x, Zenition 10 to 90, Interventional Workspot, Coronary Tools 3.1, EPIQ and Affiniti (V4.0 to 10.0), Sparq, ClearVue, CX50, Compact 5000, Echonav, IntelliSpace Portal v12, IntelliSpace Perinatal, IntelliSpace ECG, IntelliVue XDS, IntelliVue Guardian Software, Holter SW, Corsium and two ECI modules.
The flaw sits in the Windows TCP/IP stack and needs no login, which is why it rates critical. Interventional and ultrasound systems are rarely patched on the Patch Tuesday cadence. Philips policy allows only Philips authorized changes on these systems, so the patch reaches the device through Philips service.
Where IPv6 is not used on the clinical segment, turning it off at the network shrinks the window while the validated patch is scheduled.