CISA published advisory ICSMA-18-226-01 on August 14, 2018, on Philips IntelliSpace Cardiovascular (ISCV) 3.1 and earlier and Xcelera 4.1 and earlier. Philips reported the issue. It is CVE-2018-14789 (CWE-269), with a CVSS v3 base score of 4.2.
For ISCV 2.x or prior and Xcelera 4.1 or prior, an attacker with escalated privileges could access folders containing executables where authenticated users have write permissions. From there the attacker could run arbitrary code with local administrative permissions on the ISCV/Xcelera server.
Writable program folders are a classic swap point. Replace a binary that a privileged service or administrator later runs, and the replacement inherits those rights. The CVSS vector keeps the score low: local access, high privileges required and low impact on confidentiality, integrity and availability.
Philips mitigated the issue for ISCV 2.x in ISCV 3.1 and told users to contact their local Philips service support team for upgrade instructions. For ISCV 3.1 or prior and Xcelera 4.1 or prior, the fix was to arrive in ISCV 3.2, scheduled for October 2018, with interim mitigations recommended until then.
Install directory permissions are part of the product, and a cardiology server should ship with them already locked.