CISA published advisory ICSMA-17-318-01 on November 14, 2017, covering a stored credential flaw in Philips IntelliSpace Cardiovascular and Xcelera. Affected versions are IntelliSpace Cardiovascular 2.3.0 and earlier and Xcelera R4.1L1 and earlier. Philips reported the vulnerability itself.

CVE-2017-14111 (CWE-522, insufficiently protected credentials) carries a CVSS v3 base score of 7.2. Credentials sit in cleartext in system files.

The vector explains the score. Exploitation needs high privileges (PR:H), but it is network reachable with low complexity, and the confidentiality, integrity and availability impacts are all high. An attacker who already holds an elevated account can read the stored credentials and reach patient health information, system resources and connected assets.

Both products are cardiac image and information management systems, so the stored credentials are a path from one compromised account into the wider cardiology workflow. Cleartext secrets on disk turn a contained intrusion into lateral movement.

Philips said it was producing software hotfix updates for all IntelliSpace Cardiovascular and latest Xcelera versions. Some were available on request, and the rest were expected by the end of 2017. Philips also opened a voluntary medical device correction aligned with field change order FCO83000202, issued as the IntelliSpace Cardiovascular updates became available.

A self-reported flaw with a field change order attached is how coordinated disclosure is supposed to look. Credentials on disk still belong in a protected store, hashed or encrypted, from day one.