CISA published advisory ICSMA-21-322-01 on November 18, 2021, covering two vulnerabilities in the Philips IntelliBridge EC 40 and EC 80 hubs, versions C.00.04 and prior. Younes Dragoni, Andrea Palanca and Ivan Speziale of Nozomi Networks reported them.

CVE-2021-32993 (CWE-798) is hard-coded credentials, such as a password or cryptographic key, used for inbound authentication, outbound communication or encryption of internal data. CVE-2021-33017 (CWE-288) is an alternate path: the standard access path requires authentication, but another channel into the hub does not. Both score 8.1 under CVSS v3, with an adjacent network vector, no privileges and high integrity and availability impact.

An attacker who gets in could execute software, modify system configuration, or view and update files, including unidentifiable patient data. Philips notes the hub only transfers medical device data from one format to another, without controlling or altering connected devices, and is not intended for active patient monitoring. If data stops arriving from the hub, the PIC iX Central Station raises a notification and users are told to refer to the source device.

Philips planned a new release to remediate both flaws by the end of Q4 2021 and recommended interim mitigations until then. The same hubs drew CISA advisory ICSMA-19-318-01 in November 2019 for weak SSH ciphers, so this is the second round of access fixes for the product line.

A hub with one authenticated door and one open door has no authentication, so inventory every listening service before release.