CISA published advisory ICSMA-19-241-02 on August 29, 2019, covering the Philips HDI 4000 Ultrasound System running on old, unsupported operating systems such as Windows 2000. Check Point reported the issue.
CVE-2019-10988 is use of an obsolete function (CWE-477) with a CVSS v3 base score of 3.0. The finding is structural. The system is built on an operating system that no longer receives support, so any unmitigated vulnerability in that operating system could be exploited against the product. The vector requires local access, high complexity and high privileges, and the advisory notes a high skill level is needed.
CISA's risk evaluation is specific about the stakes. Exploitation could expose ultrasound images and compromise image integrity, while patient safety and system availability are not affected. Tampered images are a diagnostic problem even when the machine keeps running.
There is no patch. Philips ended the support life cycle for the HDI 4000 on December 31, 2013, and the systems are no longer sold, updated or supported. The guidance is to limit network access and consider replacing the system with newer technology on a supported operating system. A device whose operating system went out of support years earlier belongs on an isolated segment until it is replaced.