CISA published advisory ICSMA-19-113-01 on April 23, 2019, covering two vulnerabilities in Fujifilm computed radiography cassette readers: the FCR Capsula X, FCR Carbon X and FCR XC-2, all CR-IR 357. Marc Ruef and Rocco Gagliardi of Scip AG reported them.

CVE-2019-10950 (CWE-284, CVSS v3 9.8) is the serious one. The device provides telnet services with no authentication, and an attacker who reaches them may get to the underlying operating system and run arbitrary code. CVE-2019-10948 (CWE-400, CVSS 7.5) is a denial of service: an overflow of TCP packets leaves the reader needing a manual reboot, with potential image loss.

Fujifilm did not ship a patch. Its answer is a configuration called Secure Host, which tells the CR-IR 357 to ignore all network traffic except from the IP address of the Fujifilm image acquisition console. The catch is that it prevents more than one console from sharing the reader unit, so sites that share readers have to call Fujifilm to discuss options.

Fujifilm also put network security first: keep the reader on a segment only authorized devices and staff can reach, and separate guest or public traffic with a VLAN.

An open telnet port on an imaging peripheral is an admin console for anyone on the subnet, and a source IP allowlist is the least you can put in front of it.