CISA published advisory ICSMA-17-332-01 on November 28, 2017, for the Ethicon Endo-Surgery Generator Gen11. Johnson & Johnson, Ethicon's parent company, reported the flaw itself.
CVE-2017-14018 is improper authentication (CWE-287) with a CVSS v3 base score of 4.8. The authentication mechanism between the generator and its single-patient use products can be bypassed, so unauthorized devices can be connected to the generator. CISA rates the integrity impact high and availability impact low.
The vector string requires physical access and high attack complexity. An attacker has to be in the operating room with a counterfeit or modified instrument, which limits remote abuse and leaves supply chain substitution as the realistic scenario.
Ethicon contacted users and started a field cybersecurity update, made available on November 29, 2017, the day after the advisory. Questions go to the Ethicon sales representative or the Ethicon Customer Support Center.
Accessory authentication on an energy device is a safety function. If the generator cannot tell a genuine instrument from a substitute, the energy settings it applies rest on a guess.