CISA published advisory ICSMA-22-244-01 on September 1, 2022, covering five vulnerabilities in the Contec Health CMS8000 ICU/CCU vital signs patient monitor. Affected firmware is CMS7.820.075.08/0.740.75 and earlier, CMS7.820.120.01/0.930.95 and earlier, and smart3250-2.6.27-wlan2.1.7.cramfs and earlier. Level Nine reported them.

The highest score goes to CVE-2022-38100 (CWE-770, CVSS v3.1 7.5). A specially formatted UDP request from anyone with network access crashes the monitor and forces a physical reboot. Sent as a UDP broadcast, the same request can take down every CMS8000 on the network at once.

CVE-2022-36385 (CWE-1263, 6.8) needs only momentary physical access: plug in a USB drive and perform a malicious firmware update that permanently changes device functionality. No authentication or controls stand in the way, and CISA describes it as a drive-by attack that works on any CMS8000.

CVE-2022-38069 (CWE-798, 4.3) is a set of globally default credentials shared by all CMS8000 devices, which give privileged access, patient data extraction and parameter changes to anyone with momentary physical access. CVE-2022-3027 (CWE-419, 5.7) lets a Wi-Fi access point with a malicious SSID name write arbitrary files or display incorrect information when the monitor tries to connect. CVE-2022-38453 (CWE-489, 3.0) covers binaries compiled not stripped and with debug info, which makes reverse engineering easier.

No vendor fix is listed. Per FDA recommendation, CISA recommends that users remove any Contec CMS8000 devices from their networks.

A broadcast packet that crashes every monitor on a segment turns one compromised laptop into a unit-wide monitoring outage.