CISA published advisory ICSMA-17-017-02 (Update B) on January 17, 2017, covering two vulnerabilities in the Becton, Dickinson and Company (BD) Alaris 8015 PC unit, the point-of-care controller of the Alaris infusion system. The listed versions are 9.5 and prior, 9.33 and prior, and 9.7. CISA credits Dan Regalado, Asher Davila Loranca and Ryan Dion of Palo Alto.

CVE-2016-8375 (CWE-522, CVSS v3 4.9) lets an attacker with physical access remove the external Wi-Fi card and connect a pre-programmed malicious CompactFlash card to pull unencrypted wireless network authentication credentials. CVE-2016-9355 (CWE-284, CVSS v3 6.8) goes after the removable flash memory itself. An attacker who disassembles the unit gets read and write access to device memory and can overwrite application and internal data, including logs and the drug library.

The bigger exposure sits outside the pump. Software 9.5 and earlier keeps the hospital's wireless authentication credentials on that removable card, so one stolen or borrowed PC unit hands an attacker the keys to the clinical Wi-Fi network. CISA's own risk note frames the impact as access to the host facility's wireless network credentials.

BD has not developed a product fix. It issued compensating controls and a security bulletin for the 8015, and states that all Alaris System software versions below 9.19 are end of life. BD recommends upgrading when its next software version clears 510(k).

A credential stored in the clear on removable media is a facility credential, and it should be rotated and scoped as if every pump in storage could leak it.