B. Braun confirmed on October 30 that its APEX Compounding System and PINNACLE TPN Management System carry CVE-2026-109618 and CVE-2026-109619, a pair of flaws known as PHP2Shell, with a CVSS v3 base score of 7.7 under CWE-1395, dependency on a vulnerable third party component. The products use the third party package PHP, and B. Braun says the issue allows an attacker to infiltrate the product via the network.
A successful exploit could let an attacker escalate privileges, view sensitive information, upload malicious files, and perform remote code execution, and B. Braun says the resulting code execution could alter control panel software behavior, information integrity, or availability. The company says internal product surveillance discovered the third party vulnerability and that it has received no reports of exploitation or incidents in the field.
The affected list covers APEX Compounding System software 151715 (SCDM) and 151717 (TPN Manager) at revisions E (UI version 2.1.17), F (3.0.3) and G (3.0.4), and PINNACLE TPN Management System software 151706 (TPN Manager) revision C (UI version 2.0.6) and 151703 (SCDM) revision D (UI version 2.0.6). B. Braun says versions earlier than those listed are also affected, and that other components of its compounding systems are not.
B. Braun recommends installing patches that resolve the issue but has not published a fixed version number in its public statement. Until a patch is in place, the company's mitigation list calls for network segmentation and access controls, keeping compounding systems off direct internet exposure, and layering in firewalls, VLANs, monitoring or intrusion detection, none of which replace the software update itself.
PHP2Shell has driven a long run of advisories across the medical device industry this fall, most of them vendors reporting their products are not affected, including Philips, Siemens Healthineers, Fresenius Kabi and STERIS. B. Braun's APEX and PINNACLE disclosure is one of the few confirmed hits rather than a clearance, and the company followed on November 13 with two more PHP2Shell statements covering its SpaceCom platform paired with the Datamodule compactplus and with the Space Battery SP with WiFi.
A TPN compounding system is exactly the case B. Braun's own mitigation list is written for. These systems increasingly sit on the same subnet as EMR order integration and drug library feeds, and segmentation is the control standing between a network exploitable RCE and a compounding order queue. No CISA ICSMA advisory has been published for either CVE as of November 21, so pharmacy and biomed teams checking the federal feed alone will miss this one.