CISA published advisory ICSMA-22-298-01 on October 25, 2022, covering two vulnerabilities in AliveCor KardiaMobile: the Kardia App for Android, versions 5.17.1-754993421 and earlier, and all versions of the KardiaMobile device. Carlos Cilleruelo Rodríguez and Javier Junquera Sánchez of the University of Alcalá reported them.

CVE-2022-41627 (CWE-311, CVSS v3 4.8) is the design issue. The KardiaMobile device sends its readings to the phone as sound, and that data-over-sound protocol has no encryption. An attacker within 5 feet could read patient EKG results or disrupt the reading by emitting sounds at similar frequencies.

CVE-2022-40703 (CWE-302, CVSS 5.2) is in the Android app, which is open to the publicly known Intent Manipulation exploit. It lets an attacker bypass app authentication and view or alter information in the app.

AliveCor acknowledged the missing encryption but called the circumstances needed to exploit it unlikely. For the app, it points to its usage instructions, which recommend a phone passcode or biometric lock to reduce Intent Manipulation risk.

Proximity requirements lower the odds, and they do nothing about integrity: a faked cardiogram is a clinical problem no matter how close the attacker had to stand.